GE Logiq9彩超工作原理及故障案例分析

技术交流

[摘要] GE Logiq9彩色超声诊断系统有卓越的图像质量、数据分析能力及更高的工作效率,随着设备老化,在使用中会出现一些常见的故障和问题。通过介绍Logiq9彩超的结构组成及工作原理,列举由计算机(操作系统和硬件)及前端电路引起的有代表性的故障案例,阐述分析故障的思路及处理方法,总结相关经验供同行参考。

[关键词] 彩色超声诊断系统;计算机;操作系统;图像黑影DOI: 10.3969/J.ISSN.1672-8270.2013.04.032

Working principles and troubleshootings of GE Logiq9 ultrasound system/HUANG Qing-feng// China Medical Equipment,2013,10(4):87-88.

[Abstract] GE LOGIQ9 ultrasound machine has excellent image quality, powerful ability of data analysis and high work efficiency. With aging, some troubles and problems might occur. In this paper, we introduced the structure and the working principles of this machine. Then, we listed several typical troubles caused by the computer ( including the operating system and hardware system) and front-end circuit. The causes of the troubles from the starting programs and the shadows in the images were analyzed. Finally, we described the handling ways and summarized the experience of analysis strategy and handling the same kind of troubles for the reference.

[Key words] Ultrasound machine; Computer; Operating system; Image shadow

[First-author's address] Department of Equipment Management, the First Affiliated Hospital, Guangxi Medical University, Nanning 530027, China.

作者简介

黄庆丰,男,(1972- ),本科学历,工程师。广西医科大学第一附属医院设备科,从事医院医疗设备维修和管理工作。

GE Logiq9彩色超声诊断系统具有卓越的图像质量、数据分析能力及更高的工作效率,广泛应用于临床腹部、心脏、妇产、血管、泌尿系统及小器官等诊断检查,为其提供优异图像品质,综合了所有的用于患者诊断的临床表现能力。然而,随着设备的老化,在使用中会碰到一些常见的故障和问题,通过介绍GE Logiq9彩超的结构组成和工作原理,将在设备的使用中遇到的故障现象和处理方法进行分析,以供参考。

GE Logiq9彩超由电源模块、计算机(PC)、前端电路板箱、显示器和控制面板等组成。①电源模块(DC)的功能是将市电220 V交流电转变为主机各部分所需的各组工作电压;②计算机装载有Window XP操作系统(OS)和相应的应用程序,是整机的心脏,控制和协调主机各部分的工作,并对前端送来的信号进行分析计算,形成视频信号输出至显示器,得到所看到的超声图像;③前端电路由RF AMP板、TD板、EQ板及BMP板组成,主要功能是驱动超声探头晶片的起振发射超声波,并接收超声波反射信号,进行叠加及对应模式的信号处理,最后形成射频信号送至计算机进行分析处理[1-2]。1 故障案例一1.1 故障现象

机器无法启动。当机器接通市电后便听见计算机端ATX风扇运行,开机指示灯变绿。按开机键开机后指示灯变黄,风扇停转;再按开机键却无法启动,开机指示灯依然为黄色。在此过程中显示器未出现任何

①广西医科大学第一附属医院设备科 广西 南宁 530027

字符显示。

1.2 故障分析与排除

GE Logiq9彩超的电源分PC的ATX电源和主机的DC电源模块两部分,其两电源为独立分开且非同时上电,具体工作流程为:①机器接通市电后马上供电给PC的ATX电源,当按下开机键后计算机启动;②加载硬盘上的系统软件后机器启动,并在显示器显示出相应信息;③计算机内的PC2IP卡得电后发出power on信号给主机的DC电源使DC电源启动,供给各个电路板所需要的直流电源,以保证在PC软件启动的同时,各电路板到达正常工作状态。

GE Logiq9彩超机器启动首先是计算机的硬件启动,硬件启动后给出信号到主机激励DC电源供电,现机器启动的最初步骤未成功,首先应检查计算机部分。打开计算机端机箱发现,计算机端采用Intel865芯片组的主板,由ATX电源直接给计算机主板供电。因机器无法启动,应分别测试ATX电源与计算机主板的情况正常与否,以确定故障部位。此故障中检查ATX电源空载时发现各电压值均为正常,但却不能带动负载。拆开ATX电源后发现12 V电压的6个滤波电容(3300 Uf/25 V)已全部漏液,全部更换同一规格电容修复ATX电源后机器能正常启动[3-4]。1.3 小结

现代的彩超诊断仪无论何种品牌与何种型号,其工作原理为:计算机加上探头的扫描仪的结合,倘

ZHONGGUO YIXUEZHUANGBEI 87

0x0000001E**”……,蓝屏提示的代码不定,有时强8块通道板对应不同的探头通道,每个板处理24个通行关机重新启动后能正常使用,有时却无效,机器在道。在显示器图像上标记好对应的暗线位置,关机,使用过程中亦出现蓝屏死机现象。交换通道板的位置,当交换到第4块通道板时,图像2.2 故障分析与排除上对应的暗线位置发生移动,再用另一块通道板与第

机器计算机部分的软件平台是window,蓝屏4块通道板交叉检查,判断故障出现在第4块通道板,报错主要与window软件和其所依赖的计算机硬件有更换此通道板后机器恢复正常[6]。关;与超声形成相关的电路板如出故障,通常会出现3.3 小结专门的报错代码,而且不会显示蓝屏的底色。针对此现代高端彩超的信号处理通道板通常有2块、4块故障现象,首先应考虑是计算机内部出的问题,即系甚至8块,这些通道板完全相同,当出现图像有黑影统软件出错,或硬盘损坏。排除此故障的方法为:拆现象时,可调换通道板之间的位置,采用替换法达到开硬盘,用效率源软件检查硬盘可发现硬盘有小部分诊断的目的,准确锁定故障部件。坏道;更换硬盘,并重装系统后机器可正常使用。参考文献 2.3 小结[1] 赵武.阿洛卡SSD-2000超声仪故障检修分析[J].中

GE Logiq9彩超计算机内硬盘中装载有操作系统国医学装备,2008,5(5):64-66.

[2] 容丹辉.B 超原理分析及故障维修[J ].医疗装和应用软件,最重要且最基本的操作系统(OS)控制所

备,2008,21(5):50-51.有计算机运行的程序,并管理整个计算机的资源,是

[3] 刘晓华,许锋,田金.IU22彩色超声诊断仪故障维修

计算机裸机与应用程序及用户之间的桥梁。如果软件三例[J].中国医疗设备,2012,27(7):144,149.损坏或崩溃机器将无法运行。[4] 邹德春,高玉霞.彩色超声设备常见故障和判断解决3 故障案例三3.1 故障现象

超声图像出现暗线。GE Logiq9彩超的超声图像上有1条暗线,暗线区域内无回波信号,形成图像时移动探头的位置,而暗线的位置始终固定不变。

方法[J].中国医疗设备,2009,24(4):125-126.

[5] 李朝伟,王振洲,姜树勋.飞利浦IU22彩超维修故障

两例[J].中国医学装备,2011,8(6):63-64.

[6] 徐晓锋.S2000彩超硬件原理及分析[J].中国医疗设

备,2012,27(6):123-124.

收稿日期:2012-11-20

第四届全军放射物理技术专业分会学术会议召开

88 ZHONGGUO

YIXUEZHUANGBEI

技术交流

[摘要] GE Logiq9彩色超声诊断系统有卓越的图像质量、数据分析能力及更高的工作效率,随着设备老化,在使用中会出现一些常见的故障和问题。通过介绍Logiq9彩超的结构组成及工作原理,列举由计算机(操作系统和硬件)及前端电路引起的有代表性的故障案例,阐述分析故障的思路及处理方法,总结相关经验供同行参考。

[关键词] 彩色超声诊断系统;计算机;操作系统;图像黑影DOI: 10.3969/J.ISSN.1672-8270.2013.04.032

Working principles and troubleshootings of GE Logiq9 ultrasound system/HUANG Qing-feng// China Medical Equipment,2013,10(4):87-88.

[Abstract] GE LOGIQ9 ultrasound machine has excellent image quality, powerful ability of data analysis and high work efficiency. With aging, some troubles and problems might occur. In this paper, we introduced the structure and the working principles of this machine. Then, we listed several typical troubles caused by the computer ( including the operating system and hardware system) and front-end circuit. The causes of the troubles from the starting programs and the shadows in the images were analyzed. Finally, we described the handling ways and summarized the experience of analysis strategy and handling the same kind of troubles for the reference.

[Key words] Ultrasound machine; Computer; Operating system; Image shadow

[First-author's address] Department of Equipment Management, the First Affiliated Hospital, Guangxi Medical University, Nanning 530027, China.

作者简介

黄庆丰,男,(1972- ),本科学历,工程师。广西医科大学第一附属医院设备科,从事医院医疗设备维修和管理工作。

GE Logiq9彩色超声诊断系统具有卓越的图像质量、数据分析能力及更高的工作效率,广泛应用于临床腹部、心脏、妇产、血管、泌尿系统及小器官等诊断检查,为其提供优异图像品质,综合了所有的用于患者诊断的临床表现能力。然而,随着设备的老化,在使用中会碰到一些常见的故障和问题,通过介绍GE Logiq9彩超的结构组成和工作原理,将在设备的使用中遇到的故障现象和处理方法进行分析,以供参考。

GE Logiq9彩超由电源模块、计算机(PC)、前端电路板箱、显示器和控制面板等组成。①电源模块(DC)的功能是将市电220 V交流电转变为主机各部分所需的各组工作电压;②计算机装载有Window XP操作系统(OS)和相应的应用程序,是整机的心脏,控制和协调主机各部分的工作,并对前端送来的信号进行分析计算,形成视频信号输出至显示器,得到所看到的超声图像;③前端电路由RF AMP板、TD板、EQ板及BMP板组成,主要功能是驱动超声探头晶片的起振发射超声波,并接收超声波反射信号,进行叠加及对应模式的信号处理,最后形成射频信号送至计算机进行分析处理[1-2]。1 故障案例一1.1 故障现象

机器无法启动。当机器接通市电后便听见计算机端ATX风扇运行,开机指示灯变绿。按开机键开机后指示灯变黄,风扇停转;再按开机键却无法启动,开机指示灯依然为黄色。在此过程中显示器未出现任何

①广西医科大学第一附属医院设备科 广西 南宁 530027

字符显示。

1.2 故障分析与排除

GE Logiq9彩超的电源分PC的ATX电源和主机的DC电源模块两部分,其两电源为独立分开且非同时上电,具体工作流程为:①机器接通市电后马上供电给PC的ATX电源,当按下开机键后计算机启动;②加载硬盘上的系统软件后机器启动,并在显示器显示出相应信息;③计算机内的PC2IP卡得电后发出power on信号给主机的DC电源使DC电源启动,供给各个电路板所需要的直流电源,以保证在PC软件启动的同时,各电路板到达正常工作状态。

GE Logiq9彩超机器启动首先是计算机的硬件启动,硬件启动后给出信号到主机激励DC电源供电,现机器启动的最初步骤未成功,首先应检查计算机部分。打开计算机端机箱发现,计算机端采用Intel865芯片组的主板,由ATX电源直接给计算机主板供电。因机器无法启动,应分别测试ATX电源与计算机主板的情况正常与否,以确定故障部位。此故障中检查ATX电源空载时发现各电压值均为正常,但却不能带动负载。拆开ATX电源后发现12 V电压的6个滤波电容(3300 Uf/25 V)已全部漏液,全部更换同一规格电容修复ATX电源后机器能正常启动[3-4]。1.3 小结

现代的彩超诊断仪无论何种品牌与何种型号,其工作原理为:计算机加上探头的扫描仪的结合,倘

ZHONGGUO YIXUEZHUANGBEI 87

0x0000001E**”……,蓝屏提示的代码不定,有时强8块通道板对应不同的探头通道,每个板处理24个通行关机重新启动后能正常使用,有时却无效,机器在道。在显示器图像上标记好对应的暗线位置,关机,使用过程中亦出现蓝屏死机现象。交换通道板的位置,当交换到第4块通道板时,图像2.2 故障分析与排除上对应的暗线位置发生移动,再用另一块通道板与第

机器计算机部分的软件平台是window,蓝屏4块通道板交叉检查,判断故障出现在第4块通道板,报错主要与window软件和其所依赖的计算机硬件有更换此通道板后机器恢复正常[6]。关;与超声形成相关的电路板如出故障,通常会出现3.3 小结专门的报错代码,而且不会显示蓝屏的底色。针对此现代高端彩超的信号处理通道板通常有2块、4块故障现象,首先应考虑是计算机内部出的问题,即系甚至8块,这些通道板完全相同,当出现图像有黑影统软件出错,或硬盘损坏。排除此故障的方法为:拆现象时,可调换通道板之间的位置,采用替换法达到开硬盘,用效率源软件检查硬盘可发现硬盘有小部分诊断的目的,准确锁定故障部件。坏道;更换硬盘,并重装系统后机器可正常使用。参考文献 2.3 小结[1] 赵武.阿洛卡SSD-2000超声仪故障检修分析[J].中

GE Logiq9彩超计算机内硬盘中装载有操作系统国医学装备,2008,5(5):64-66.

[2] 容丹辉.B 超原理分析及故障维修[J ].医疗装和应用软件,最重要且最基本的操作系统(OS)控制所

备,2008,21(5):50-51.有计算机运行的程序,并管理整个计算机的资源,是

[3] 刘晓华,许锋,田金.IU22彩色超声诊断仪故障维修

计算机裸机与应用程序及用户之间的桥梁。如果软件三例[J].中国医疗设备,2012,27(7):144,149.损坏或崩溃机器将无法运行。[4] 邹德春,高玉霞.彩色超声设备常见故障和判断解决3 故障案例三3.1 故障现象

超声图像出现暗线。GE Logiq9彩超的超声图像上有1条暗线,暗线区域内无回波信号,形成图像时移动探头的位置,而暗线的位置始终固定不变。

方法[J].中国医疗设备,2009,24(4):125-126.

[5] 李朝伟,王振洲,姜树勋.飞利浦IU22彩超维修故障

两例[J].中国医学装备,2011,8(6):63-64.

[6] 徐晓锋.S2000彩超硬件原理及分析[J].中国医疗设

备,2012,27(6):123-124.

收稿日期:2012-11-20

第四届全军放射物理技术专业分会学术会议召开

88 ZHONGGUO

YIXUEZHUANGBEI


相关内容

  • 超声诊断仪维修
  • GE Vivid Five超声诊断仪图像故障维修 Vivid Five是GE 公司在2002年推出的一款高档全数字彩色多普勒超声诊断系统,该机具有优异的图像质量和强大的处理功能,具有无创.简便.快速诊断的优点,对各种复杂疾病的诊断尤其是心血管疾病诊断具有重要辅助作用.该机除可进行消化科.泌尿科.妇产 ...

  • [推荐案例]丰田皇冠3.0L轿车电控发动机热启动困难的维修
  • 车辆信息: 一辆皇冠3.0L轿车,发动机型号为2JZ-GE.冷启动正常,当发动机运行一段时间,热态下启动发动机,则须启动多次才能着车.车主还反映:该车近段时间油耗比以前大.人工调取故障码,显示电控系统无故障码存储. 故障分析: 根据汽油发动机的工作原理及该车在启动时的控制原理分析,造成冷车启动正常, ...

  • 变压器保护系统
  • 变压器保护系统 性能价格优良的可接入 6绕组变压器保护 主要优点 独特的变压器保护功能,监视模拟和RTD输入 通过双原理(差动和方向)以及CT饱和检测的应用提高安全性-同类专利解决方案中原理一流,只需要2ms不受到饱和影响的电流即可做出正确判断 通过紧凑的设计减少安装所需空间-多功能装置集成保护和控 ...

  • 培 训 大 纲
  • 中国电信2007年度 宽带业务维护服务技能竞赛 培 训 大 纲 集团公司竞赛办公室 二〇〇七年五月 前 言 <中国电信2007年度宽带业务维护服务技能竞赛培训大纲>由基础类培训大纲.终端类培训大纲.接入类培训大纲.IP 城域骨干网培训大纲和IP 长途网培训大纲等五部分组成,每部分包括课程 ...

  • 辐射安全与防护状况2016年度评估报告
  • 章丘区中医医院2016年度辐射安全与 防护状况年度评估报告 (报告编写单位名称并盖章) 二O 一六 年 十二 月 目 录 一.医院基本情况····································1 二.放射性同位素与射线装置台帐······················ 三.辐射环 ...

  • 实习报告医院
  • 下午1点15分,我们生物医学工程2000级全体同学,共30名,在学校大门集合完毕.1点30分,我们在张老师的带领下乘车前往本次实习的地点,重庆市中山医院. 在路途中,张老师为我们讲解了本次实习的目的及重点,并且提出了一些参观实习中需要注意的纪律和要求. 我们生物医学工程专业所学习的重点在于各种医疗仪 ...

  • 中山医院参观实习总结
  • 参观时间: 2005年2月18日 参观地点: 重庆市中山医院 带队老师: 王平 2月18日,星期三,天气晴好。 下午1点15分,我们生物医学工程2000级全体同学,共30名,在学校大门集合完毕。1点30分,我们在王平老师的带领下乘车前往本次实习的地点,重庆市中山医院。 在路途中,王平老师为我们讲解了 ...

  • 参观实习总结
  • 参观时间: 2005年2月18日 参观地点: 重庆市中山医院 带队老师: 王平 2月18日,星期三,天气晴好。 下午1点15分,我们生物医学工程2000级全体同学,共30名,在学校大门集合完毕。1点30分,我们在王平老师的带领下乘车前往本次实习的地点,重庆市中山医院。 在路途中,王平老师为我们讲解了 ...

  • 医院实习报告
  • 下午1点15分,我们生物医学工程xx级全体同学,共30名,在学校大门集合完毕。1点30分,我们在张老师的带领下乘车前往本次实习的地点,重庆市中山医院。 在路途中,张老师为我们讲解了本次实习的目的及重点,并且提出了一些参观实习中需要注意的纪律和要求。 我们生物医学工程专业所学习的重点在于各种医疗仪器, ...